TITLE

Design and calibration of a fast-time resolution charge exchange analyzer

AUTHOR(S)
Scime, Earl; Hokin, Samuel
PUB. DATE
October 1992
SOURCE
Review of Scientific Instruments;Oct92, Vol. 63 Issue 10, p4527
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
A five channel, fast time resolution, scanning charge exchange analyzer has been developed for the Madison Symmetric Torus (MST). The analyzer consists of an iron vacuum vessel, a gas stripping cell, an electrostatic bending field, and five continuous electron multiplier detectors. The incident neutral flux and operation of the detectors in current mode limits the time resolution of the analyzer to 10 µs. The analyzer was absolutely calibrated over the energy range of interest (500-2000 eV) with a H[sup +] beam, so that the charge exchange power loss could also be measured. The analyzer can be swiveled on a shot-to-shot basis for measurements of T[sub i](r), where 0.3
ACCESSION #
9787117

 

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.