TITLE

A solid dielectric Compton diode for measuring short radiation pulse widths

AUTHOR(S)
Carlson, G. A.; Sanford, T. W. L.; Davis, B. A.
PUB. DATE
November 1990
SOURCE
Review of Scientific Instruments;Nov90, Vol. 61 Issue 11, p3447
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
A simple, rugged, and inexpensive solid dielectric Compton diode (SCD) has been developed for measuring the radiation-time history of flash x-ray sources. By limiting the physical size of the Compton diode and matching its characteristic impedance to that of the signal cable, a frequency response above 1 GHz has been achieved. The SCD has been used to measure the pulse width and rise time of HERMES III, a pulsed γ-ray simulator. A second, prototype SCD has been built for use with Saturn, a pulsed x-ray simulator. The SCDs do not have high sensitivity, and are thus limited to measurements of dose rate above 109 rads/s, due to signal-to-noise limitations. At lower dose rates, more sensitive detectors such as scintillator photodiodes or p-i-n diodes are more suitable. On the other hand, for dose rates above 1010 rads/s, the more sensitive detectors tend to saturate, while the SCD continues to respond linearly to at least 1012 rads/s. Thus, the SCD is well suited for measuring radiation-time histories at high dose rates on flash x-ray sources.
ACCESSION #
9786132

 

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library