TITLE

Absolute calibration of a SPRED EUV spectrograph for use on the DIII-D tokamak (abstract)

AUTHOR(S)
Wood, R. D.; Allen, S. L.
PUB. DATE
August 1988
SOURCE
Review of Scientific Instruments;Aug1988, Vol. 59 Issue 8, p1537
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
We have performed an absolute intensity calibration of a SPRED multichannel EUV spectrograph using synchrotron radiation from the NBS SURF-II electron storage ring. The calibration procedure and results for both a survey (450 g/mm) and a high-resolution (2100 g/ mm) grating are presented. The spectrograph is currently in use on the DIII-D tokamak with a tangential line-of-sight at the plasma midplane. Data are first acquired and processed by a microcomputer; the absolute line intensities are then sent to the DIII-D database for comparison with data from other diagnostics. Representative absolute line brightnesses for ohmic, L-mode, and H-mode plasma operations will be presented. This work was supported by the U.S. DOE Contracts W-7405-ENG-48 and DE-AC03-84ER51044.
ACCESSION #
9781464

 

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.