TITLE

First measurement of neutron emission profile on JT-60U using Stilbene neutron detector with neutron-gamma discrimination

AUTHOR(S)
Ishikawa, M.; Nishitani, T.; Morioka, A.; Takechi, M.; Shinohara, K.; Shimada, M.; Miura, Y.; Nagami, M.; Kaschuck, Yu. A.
PUB. DATE
December 2002
SOURCE
Review of Scientific Instruments;Dec2002, Vol. 73 Issue 12, p4237
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
This article describes a diagnostic for measuring neutron emission profile in JT-60U. The Stilbene neutron detector, developed by TRINITI laboratory in Russia, has been installed on the JT-60U Tokamak to measure the neutron emission profile for the first time. The Stilbene neutron detector is a detector which combines a Stilbene crystal scintillator with a neutron-gamma pulse shape discrimination circuit, with a very compact size. Performance tests were carried out using neutron and gamma-ray source prior to installation on JT-60U. Good gamma suppression of the Stilbene neutron detector was verified. Though the neutron emission profile obtained by Stilbene neutron detectors has error of 30% in innermost channel with a calculation using measured plasma parameters, there is an agreement within 10% error in the other channels.
ACCESSION #
8563473

 

Related Articles

  • Conceptual design for a scintillating-fiber neutron detector for fusion reactor plasma diagnostics. Sailor, W.C.; Barnes, Cris W.; Chrien, R.E.; Wurden, G.A. // Review of Scientific Instruments;Jan1995, Vol. 66 Issue 1, p898 

    Describes a conceptual design for pointing neutron detector capable of delivering count rates of events from triton burnup at a deuterium-burning tokamak. Collimated bundles of scintillating bundles consisting the detector; Separation by metal and polyethylene; Bundles set into a larger collimator.

  • Long-term performance of neutron detectors during deuterium-tritium operation in the Tokamak... Jassby, D.L.; Barnes, Cris W.; Bitter, M.; Johnson, L.C.; Lemunyan, G.; Roquemore, A.L.; von Goeler, S. // Review of Scientific Instruments;Jan1999, Vol. 70 Issue 1, p1111 

    Examines the long-term performance of neutron detectors during deuterium-tritium operation in the Tokamak Fusion Test Reactor. Overview on the detector; Comparison of neutron yields; Statistics of yield ratios; Physical reliability.

  • Scintillating-fiber 14 MeV neutron detector on TFTR during DT operation. Wurden, G.A.; Chrien, R.E.; Barnes, Cris W.; Sailor, W.C.; Roquemore, A.L.; Lavelle, M.J.; O'Gara, P.M.; Jordan, R.J. // Review of Scientific Instruments;Jan1995, Vol. 66 Issue 1, p901 

    Tests a compact neutron detector using an array of scintillating fibers on the Tokamak Fusion Test Reactor (TFTR) tokamak under conditions of a high gamma background. Use of a fiber-matrix geometry; Magnetic field-insensitive phototube with an active base; Pulse-height discrimination to reject...

  • A proposed diagnostic for time-resolved 14 MeV neutron measurements on TFTR. Ku, L. P.; Nazikian, R.; Prorvitch, V. // Review of Scientific Instruments;Oct90, Vol. 61 Issue 10, p3193 

    A novel method for time-resolved measurements of the 14 MeV neutron flux in an intense 2.5 MeV neutron and γ-ray background has been developed. Discrimination against the background 2.5 MeV neutron and γ-ray flux is achieved by the use of polyethylene and lead shielding. A high detection...

  • Time-resolved 14-MeV neutron detector for triton confinement studies. Smith, J. R.; Fisher, R. K.; Leffler, J. S.; Newman, D. E. // Review of Scientific Instruments;Aug1986, Vol. 57 Issue 8, p1780 

    Previous measurements of triton burnup in deuterium tokamak plasmas show an anomalously small burnup for low-q discharges with significant magnetohydrodynamic activity. This may have important implications for alpha particle burnup in a fusion reactor. By developing a detector capable of...

  • Measurements of the neutron source strength at DIlI-D Heidbrink, W. W.; Taylor, P. L.; Phillips, J. A. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p536 

    Documents the calibration procedure for neutron counters and scintillators used to measure the neutron emission produced by the DIII-D tokamak. Detectors and calibration; Detector comparisons; Comparison with other diagnostics.

  • TFTR natural diamond detectors based D-T neutron spectrometry system Krasilnikov, A. V.; Azizov, E. A.; Roquemore, A. L.; Khrunov, V. S.; Young, K. M. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p553 

    Describes the development of three natural diamond detector-based spectrometry systems and their use on tokamak fusion test reactor to perform deuterium-tritium neutron spectrum and flux measurements. Detector operation and performance characteristics; Diagnostic arrangement; Initial plasma...

  • Determination of TFTR far-field neutron detector efficiencies by local neutron flux spectrum measurement Jassby, D. L.; Ascione, G.; Kugel, H. W.; Roquemore, A. L.; Barcelo, T. W.; Kumar, A. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p540 

    Presents the results of local neutron flux spectrum measurements conducted to evaluate the performance of neutron detectors located on the tokamak fusion test reactor. Spectral fluence from activation measurements; Tritium breeding in [sup 3]He; Surface-barrier diodes and natural diamond...

  • Calibration issues of the Tokamak Fusion Test Reactor multichannel neutron collimator. von Goeler, S.; Johnson, L. C.; Bitter, M.; Efthimion, P. C.; Roquemore, A. L. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p548 

    Evaluates the calibration procedures for the detectors in the neutron collimator of the Tokamak Fusion Test Reactor. Deuterium-tritium jog shot calibration of the Bay A detectors; Detector nonlinearity; Deuterium-deuterium (DD) jog shot for the scattered DD neutron background.