TITLE

Identification of errors in the electron density measurements of a tangential interferometer/polarimeter system during a tokamak discharge

AUTHOR(S)
Arakawa, H.; Kawano, Y.; Itami, K.
PUB. DATE
October 2012
SOURCE
Review of Scientific Instruments;Oct2012, Vol. 83 Issue 10, p10E345
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
A new method for the comparative verification of electron density measurements obtained with a tangential interferometer and a polarimeter during a discharge is proposed. The possible errors associated with the interferometer and polarimeter are classified by the time required for their identification. Based on the characteristics of the errors, the fringe shift error of the interferometer and the low-frequency noise of the polarimeter were identified and corrected for the JT-60U tangential interferometer/polarimeter system.
ACCESSION #
82964786

 

Related Articles

  • Two-color polarimeter for electron density measurement on large tokamaks. Kawano, Yasunori; Nagashima, Akira // Review of Scientific Instruments;Nov97, Vol. 68 Issue 11, p4035 

    Proposes a two-color laser polarimeter on large tokamaks for electron density measurement based on the midplane Faraday rotation. Elimination of the Faraday rotation at vacuum windows from measured polarization angles; Photoelastic modulators used for stable polarization measurement.

  • Two-wavelength infrared interferometer/polarimeter system for CIT. Ma, C. H.; Hutchinson, D. P.; Vander Sluis, K. L. // Review of Scientific Instruments;Aug1988, Vol. 59 Issue 8, p1629 

    The results of a feasibility study of a two-wavelength infrared interferometer/polarimeter system for measurements of electron density and plasma current profiles in the Compact Ignition Tokamak (CIT) are presented. The system utilizes CO[sub 2] lasers at a wavelength of 10.6 µm, and...

  • Multichannel submillimeter interferometer/polarimeter for TdeV. Lachambre, J.-L.; Gagné, M. // Review of Scientific Instruments;Nov94, Vol. 65 Issue 11, p3428 

    An interferometer/polarimeter has been developed and tested for operation on the TdeV tokamak. The system operates at 393.6 µm on 7 channels equally spaced over the plasma diameter. The interferometer has operated routinely for six years and is used for density feedback control on TdeV, with...

  • Density profile reconstructions from 2-D interferometric data on Microtor using novel tomographic analysis techniques (abstract). Howard, J.; Doyle, E. J.; Reibeiz, G.; Savage, R. L.; Peebles, W. A.; Luhmann, N. C. // Review of Scientific Instruments;Aug1988, Vol. 59 Issue 8, p1592 

    Plasma tomographic reconstructions are subject to aliasing ambiguity as a result of the limited angular and radial sampling rates for the line-integrated data. The two-view interferometer installed on the Microtor tokamak yields unambiguous information, specifically, a collection of six...

  • A multichannel interferometer for electron density measurements in COMPASS. Edlington, Trevor; Wylde, Richard // Review of Scientific Instruments;Oct92, Vol. 63 Issue 10, p4968 

    A compact seven channel interferometer has been designed and built to measure electron density profiles in the COMPASS (compact assembly) tokamak. Two far-infrared (FIR) laser cavities are optically pumped with a single continuous-wave CO[sub 2] laser, generating two similar beams at λ = 433...

  • Design of far-infrared three-wave polarimeter-interferometer system for the J-TEXT tokamak. Chen, J.; Gao, L.; Zhuang, G.; Wang, Z. J.; Gentle, K. W. // Review of Scientific Instruments;Oct2010, Vol. 81 Issue 10, p10D502 

    A multichannel three-wave far-infrared polarimeter-interferometer will be constructed on the J-TEXT tokamak (R=1.05 m, a=0.27 m, BT≤3 T, and IP≤350 kA) for current density profile and electron density profile measurements. The system will adopt the three-wave polarimeter...

  • Multichannel interferometer/polarimeter system for the RTP tokamak. van Lammeren, A. C. A. P.; Kim, S. K.; Donné, A. J. H. // Review of Scientific Instruments;Oct90, Vol. 61 Issue 10, p2882 

    A nine-channel interferometer/polarimeter system is designed for the RTP tokamak (Rijnhuizen Tokamak Project, a = 0.16 m, R = 0.72 m, B[sub T] = 2.5 T, I[sub p] ≤ 200 kA, and plasma pulse duration 200 ms). A CO[sub 2]-pumped dual-cavity FIR laser system is used to obtain two FIR laser...

  • Multichannel far-infrared polarimeter system on TEXT-Upgrade Brower, D. L.; Zeng, L.; Jiang, Y. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p419 

    Describes the addition of a far-infrared polarimeter diagnostic to the University of California at Los Angeles multichord heterodyne interferometer system on the TEXT-Upgrade tokamak. Design of the polarimeter/interferometer system; Equilibrium profiles; Tearing modes; Fast current ramps and...

  • CO2 laser interferometer for electron density measurement in JT-60U tokamak. Kawano, Yasunori; Nagashima, Akira; Ishida, Shinichi; Fukuda, Takeshi; Matoba, Tohru // Review of Scientific Instruments;Oct92, Vol. 63 Issue 10, p4971 

    A tangential chord CO[sub 2] laser interferometer for electron density measurement in the JT-60U tokamak is described. To compensate for a phase shift due to mirror vibrations, an IR-HeNe laser interferometer is introduced. Probe beams of the CO[sub 2] and the IR-HeNe lasers travel in a similar...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library