The relative safety of an oxygenator

AUTHOR(S)
Svenmarker, S.; Haggmark, S.; Jansson, E.; Lindholm, R.; Appelblad, M.; Aberg, T.
PUB. DATE
September 1997
SOURCE
Perfusion;1997, Vol. 12 Issue 5, p289
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
Analysis of prospectively registered incidents related to cardiopulmonary bypass (CPB) was initiated to establish the incident rate for 10 different oxygenator brands employed over a seven-year period in 5000 clinical perfusions. A general safety index (SI) was defined as the number of recorded incidents in a given series of oxygenators divided by its total number and multiplied by 100. Specific SI was calculated for each of the following categories: high-pressure drop, debris, gas exchange, leakage, material failure and oxygenator change-out. An SI of 0.2 was arbitrarily set as a reference and an acceptable safety level. An estimate of the relative risk for a particular oxygenator brand was compared with the reference by calculating the odd's ratio with a 95% confidence interval. The mean SI was determined to be 1.6, ranging from 0 for the Maxima CBAS and the Cobe CML to 2.92 for the Safe oxygenator. The dominating specific type of incident was HPD with an SI of 0.81 followed by debris, SI = 0.71. A systematic analysis of adverse events in CPB may be used to evaluate and to set standards, a method already employed in the pharmaceutical industry. Our results indicate that oxygenator safety margins may vary between different brands.
ACCESSION #
7393742

 

Related Articles

  • Editorial. Taylor, K. // Perfusion;1998, Vol. 13 Issue 3, p155 

    Editorial. Discusses on the ability of membrane oxygenators to handle and remove air. Quantification and localization of the air removal characteristics of membrane oxygenators; Analysis on the design and manufacture of cardiopulmonary bypass circuit components for neonatal and pediatric...

  • The incidence and cause of emergency oxygenator changeovers. Fisher, A.R. // Perfusion;1999, Vol. 14 Issue 3, p207 

    Emergency oxygenator changeover presents as an important potential hazard during cardiopulmonary bypass. In order to assess the frequency and possible causes of this event a retrospective survey was conducted of cardiothoracic units in the UK and Ireland. The survey was sent to these units in...

  • Cryofibrination: did it really happen? Luckraz, H.; Boolauky, D.; Mohangee, A. // Perfusion;Jan2001, Vol. 16 Issue 1, p83 

    Cryofibrin precipitation is a phenomenon which has only recently been described, although it might have been present since the advent of membraneoxygenators for cardiopulmonary bypass procedures. The deposition of the cold-reactive protein, cryofibrin, causes progressive occlusion of the...

  • Parallel replacement of the oxygenator that is not transferring oxygen: the PRONTO procedure. Groom, R. C.; Forest, R. J.; Cormack, J. E.; Niimi, K. S.; Morton, J. // Perfusion;Nov2002, Vol. 17 Issue 6, p447 

    We present here a technique to replace a failed oxygenator by inserting a second oxygenator in parallel (PRONTO) within the cardiopulmonary bypass (CPB) circuit. Oxygenator failure is a potential hazard that may result in patient injury or death. Although failures are rare, safety surveys...

  • Can an oxygenator design potentially contribute to air embolism in cardiopulmonary bypass? A novel method for the determination of the air removal capabilities of neonatal membrane oxygenators. de Somer, F.; Dierickx, P.; Dujardin, D.; Verdonck, P.; Nooten, G.V. // Perfusion;1998, Vol. 13 Issue 3, p157 

    At present, air handling of a membrane oxygenator is generally studied by using an ultrasonic sound bubble counter. However, this is not a quantitative method and it does not give any information on where air was entrapped in the oxygenator and if it eventually was removed through the membrane...

  • Oxygenator thrombosis: an international phenomenon. Schaadt, J. // Perfusion;1999, Vol. 14 Issue 6, p425 

    Oxygenator thrombosis, despite adequate anticoagulation, has become a recent concern amongst perfusionists worldwide. The phenomenon is characterized by a transient increase in oxygenator inlet pressure of up to 900 mmHg about 10-15 min after the institution of cardiopulmonary bypass (CPB)....

  • A clinical evaluation of platelet function, haemolysis and oxygen transfer during cardiopulmonary bypass comparing the Quantum HF-6700 to the HF-5700 hollow fibre membrane oxygenator. Chukwuemeka, AO; Turtle, M. R. J.; Trivedi, U. H.; Venn, G. E.; Chambers, D. J. // Perfusion;Oct2000, Vol. 15 Issue 6, p479 

    The continued improvement of oxygenators is an important aspect of patient safety during cardiopulmonary bypass (CPB). The purpose of this study was to compare the Bard William Harvey HF-5700 oxygenator to the upgraded Bard Quantum HF-6700, which has recently been introduced into clinical...

  • Methylene Blue Administration Mimics an Oxygenator Failure during Cardiopulmonary Bypass. Saczkowski, Richard; Smith, Michael // Internet Journal of Thoracic & Cardiovascular Surgery;2009, Vol. 13 Issue 1, p2 

    An oxygenator failure during cardiopulmonary bypass is an emergency. The entire cardiac surgical team requires a comprehensive and methodical approach to troubleshoot the possible cause. We report that the bolus administration of methylene blue during cardiopulmonary bypass will induce a...

  • Air trapping ability of the Spiral Goldâ„¢ membrane oxygenator: an ex vivo study. Mueller, Xavier M; Tevaearai, Hendrick T; van Ness, Karen; Horisberger, Judith; Augstburger, Monique; Burki, Marco; von Segesser, Ludwig K // Perfusion;1998, Vol. 13 Issue 1 

    Despite an overall improvement in cardiopulmonary bypass (CPB) technology and materials, air emboli still occur. The latest generation membrane oxygenator from Bentley Laboratories, the SpiralGoldâ„¢, was tested ex vivo for its air handling ability. The study was conducted on four calves....