TITLE

Central venous catheters may be a potential source of massive air emboli during vascular procedures involving extracorporeal circulation: an experimental study

AUTHOR(S)
Høgetveit, J.O.; Saatvedt, K.; Norum, H.; Kristiansen, F.; Elvebakk, O.; Dahle, G.; Geiran, O.R.
PUB. DATE
July 2011
SOURCE
Perfusion;Jul2011, Vol. 26 Issue 4, p341
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
Central venous catheters are mandatory during every major procedure involving extracorporeal circulation. Air emboli potentially could enter the circulation through this device when negative pressure is applied in the venous cannula. The following experimental study was initiated by a fatal massive air embolus during a vascular procedure involving cardiopulmonary bypass. An experimental setup was established, simulating a real scenario. The experiment was performed with a 40% glycerol/water mixture which exhibits properties and fluid dynamics close to blood. A heart-lung machine provided circulation of the fluid. The flow was adjusted according to the gravitational status. A triple-lumen central venous catheter with one line open to air was lowered into the liquid. The disconnected lumen of the central venous catheter was manipulated so it approached and was located in close proximity to the venous cannula. An air flow of up to 300 ml/min could be obtained from the central venous catheter with a flow in the cardiopulmonary bypass circuit of 2.3 L/min. A linear relationship was observed between flow in the circuit and air flow. Consecutive measurements proved consistent with acceptable results, proving that a disconnected central venous catheter might, under certain circumstances, be a source of massive air emboli during cardiopulmonary bypass.
ACCESSION #
62586986

 

Related Articles

  • Gaseous microemboli in cardiac surgery with cardiopulmonary bypass: the use of veno-arterial shunt as a preventive method. dos Reis, Edison Emidio; de Menezes, Livia Dutra; Lanaro Justo, Caio César // Brazilian Journal of Cardiovascular Surgery;jul-set2012, Vol. 27 Issue 3, p436 

    Neurological complications are an important cause of morbidity in the postoperative period of cardiac surgery and its incidence reaches up to 75% of patients. An important cause of these events is the formation of microbubbles in the bloodstream during cardiopulmonary bypass. Integrative review...

  • Does CO2 flushing of the empty CPB circuit decrease the number of gaseous emboli in the prime? Nyman, J.; Rundby, C.; Svenarud, P.; van der Linden, J. // Perfusion;Jul2009, Vol. 24 Issue 4, p249 

    Twenty (20) CPB-circuits were randomized to a CO2 group or a control group. In the CO2 group, each circuit was flushed with CO2 (10L/min) at the top of the venous reservoir for 5 minutes, after which priming fluid was added without interruption of the CO2 inflow. Control group circuits were not...

  • Extracorporeal circulation for port-access cardiac surgery. Toomasian, J.M.; Peters, W.S.; Siegel, L.C.; Stevens, J.H. // Perfusion;1997, Vol. 12 Issue 2, p83 

    Minimally invasive techniques for cardiac surgery are a new approach in performing some cardiac operations. Minimally invasive surgery may minimize patient discomfort, length of stay in the hospital and postoperative rehabilitation. These procedures utilize a small thoracotomy for direct...

  • Low-prime perfusion circuit and autologous priming in CABG surgery on a Jehova's Witness: a case report. van Kempen, RAB; Gasiorek, JM; Bloemendaal, K; Storm van Leeuwen, RPH; Bulder, ER // Perfusion;Jan2002, Vol. 17 Issue 1, p69 

    Cardiac surgery on Jehovah's Witnesses is a great challenge for the cardiothoracic surgery team and especially for the perfusionist. To reduce the risk of surgery in these patients, a very small extracorporeal circuit was designed to decrease the amount of priming volume and thereby the degree...

  • Modular minimally invasive extracorporeal circulation systems; can they become the standard practice for performing cardiac surgery? Anastasiadis, K; Antonitsis, P; Argiriadou, H; Deliopoulos, A; Grosomanidis, V; Tossios, P // Perfusion;Apr2015, Vol. 30 Issue 3, p195 

    Minimally invasive extracorporeal circulation (MiECC) has been developed in an attempt to integrate all advances in cardiopulmonary bypass technology in one closed circuit that shows improved biocompatibility and minimizes the systemic detrimental effects of CPB. Despite well-evidenced clinical...

  • Commentary on: The use of percutaneous ECMO support as a 'bridge to bridge' in heart failure patients: a case report. Fitzgerald, David // Perfusion;Sep2010, Vol. 25 Issue 5, p327 

    In this article, the author comments on the article "The use of percutaneous ECMO support as a 'bridge to bridge' in heart failure patients: a case report." The author says that doctors keep a fully functioning primary console as a backup along with an additional motor and a flow probe when...

  • System Makes Heart Surgery More Tolerable.  // Design News;3/15/2004, Vol. 60 Issue 4, p78 

    Features the cardiopulmonary bypass system created by CardioVention Inc. Advantages over conventional systems; Functions integrated in the solution of CardioVention; Components of the system.

  • HEART SUPPORT SYSTEMS: FROM IDEA TO CLINICAL PRACTICE. Del Cañzo, Juan F. // Contributions of Macedonian Academy of Sciences & Arts;2008, Vol. 29 Issue 2, p61 

    This paper describes the development of a pneumatic ventricular assist device by our group from the firsts steps to the clinical assay. The device, called BCM, has an input compliant cannula that improves the filling of the device. Results of the clinical assay showed that the BCM device was...

  • Use of Right Ventricular Support with a Centrifugal Pump in Post-Valve Surgery Right Ventricular Failure: A Case Series. Moulodi, Abdol Rasoul; Sheibat Zadeh, Gholam Reza; Sabzi, Feridoun // Journal of Tehran University Heart Center;2014, Vol. 9 Issue 1, p38 

    The optimal treatment method for right ventricular failure after valve surgery complicated by a low cardiac output has not been determined, although several case reports have been published on patients with ventricular failure and arrhythmia who were bridged to cardiac transplantation using...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics