TITLE

Detection of high frequency intensity oscillations at RESEDA using the CASCADE detector

AUTHOR(S)
Häussler, W.; Böni, P.; Klein, M.; Schmidt, C. J.; Schmidt, U.; Groitl, F.; Kindervater, J.
PUB. DATE
April 2011
SOURCE
Review of Scientific Instruments;Apr2011, Vol. 82 Issue 4, p045101
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
We have explored the technological potential of combining neutron resonance spin echo (NRSE) with the time-of-flight method in quasielastic neutron scattering (QENS) experiments. For these test measurements at the new NRSE instrument RESEDA (FRM II, Munich), we have employed CASCADE, one of the fastest neutron detectors in the world, developed at the University of Heidelberg. Conventionally, scintillation detectors are used, in order to detect neutron intensities with high time resolution. In contrast, we used the new CASCADE detector converting neutrons in thin 10B layers being capable of resolving neutron intensity modulations up to the megahertz regime. This fast detector allows us to abandon the last resonance flip coil of a standard NRSE setup. The classical spin echo signal is replaced by a time-modulated signal. In this setup, fast intensity modulations are present at the detector position. In order to demonstrate, that NRSE-CASCADE operates well up to detector frequencies of 10 MHz, we performed elastic polarization test measurements on a standard sample. The CASCADE detector is a multidetector accumulating counts in 128 × 128 pixels on a surface of 200 mm × 200 mm. We have analyzed the signal in 600 pixels, providing information about the spin phase reaching the detector and about the resolution function of this new variant tested at RESEDA.
ACCESSION #
60283562

 

Related Articles

  • Neutron yield measurements with current-mode neutron time-of-flight detectors (abstract). Cable, M. D.; Nelson, M. B. // Review of Scientific Instruments;Aug1988, Vol. 59 Issue 8, p1738 

    Current-mode neutron time-of-flight detectors (e.g., a scintillator and photomultiplier tube) are frequently used as sensitive detectors for laser fusion neutron yield measurements. For very low yields, as in the measurement of secondary (dt) neutron, s from deuterium targets, some subtle...

  • A new scintillation-type time-of-flight anion detector. Pramann, Axel; Rademann, Klaus // Review of Scientific Instruments;Aug2001, Vol. 72 Issue 8 

    A new scintillation-type anion time-of-flight detector of high transmittance is reported. The precursor type of this detector, which is known as the Daly detector, is often used in molecular beam experiments in combination with time-of-flight mass spectrometry. We present an improved and simple...

  • Theoretical highlights on neutrino-nucleus interactions: current challenges. Alvarez-Ruso, L. // AIP Conference Proceedings;10/6/2011, Vol. 1382 Issue 1, p161 

    The present theoretical status of neutrino interactions in the few-GeV region is reviewed. Quasielastic scattering, pion production and their importance for neutrino oscillation studies are discussed, making emphasis on the open questions that arise in the comparison with new experimental data.

  • Coherent dynamics of meta-toluidine investigated by quasielastic neutron scattering. Faraone, Antonio; Hong, Kunlun; Kneller, Larry R.; Ohl, Michael; Copley, John R. D. // Journal of Chemical Physics;3/14/2012, Vol. 136 Issue 10, p104502 

    The coherent dynamics of a typical fragile glass former, meta-toluidine, was investigated at the molecular level using quasielastic neutron scattering, with time-of-flight and neutron spin echo spectrometers. It is well known that the static structure factor of meta-toluidine shows a prepeak...

  • An experimental setup for high resolution 10.5 eV laser-based angle-resolved photoelectron spectroscopy using a time-of-flight electron analyzer. Berntsen, M. H.; Götberg, O.; Tjernberg, O. // Review of Scientific Instruments;Sep2011, Vol. 82 Issue 9, p095113 

    We present an experimental setup for laser-based angle-resolved time-of-flight photoemission. Using a picosecond pulsed laser, photons of energy 10.5 eV are generated through higher harmonic generation in xenon. The high repetition rate of the light source, variable between 0.2 and 8 MHz,...

  • Parallel Configuration For Fast Superconducting Strip Line Detectors With Very Large Area In Time Of Flight Mass Spectrometry. Casaburi, A.; Zen, N.; Suzuki, K.; Ejrnaes, M.; Pagano, S.; Cristiano, R.; Ohkubo, M. // AIP Conference Proceedings;12/16/2009, Vol. 1185 Issue 1, p80 

    We realized a \ery fast and large Superconducting Strip Line Detector based on a parallel configuration of nanowires. The detector with size 200×200 μm2 recorded a sub-nanosecond pulse width of 700 ps in FWHM (400 ps rise time and 530 ps relaxation time) for lysozyme monomers/multimers...

  • Studying transient phase dynamics using the phase resetting method. Pavlov, A. N. // Technical Physics Letters;Dec2006, Vol. 32 Issue 12, p1067 

    A new method is proposed for the investigation of transitions from synchronous dynamics of coupled oscillators to asynchronous regimes attained upon synchronization breakage by a sequence of short phase resetting pulses. The possibility to use the proposed method for determining the coefficient...

  • SENSOR FUSION AND MOTION ANALYSIS FOR MEDICAL REHABILITATION--THE FRAUNHOFER FIRST MOTION TECHNOLOGY VALIDATION LAB. KOCK, GERD; JOHN, MICHAEL; KLOSE, STEFAN; ERNST, THILO; POPESCU, FLORIN // Systems Science;2010, Vol. 36 Issue 1, p41 

    It is a well-known factor that our activities of daily life play an important role for staying healthy. In order to validate different motion analysis techniques for clinical and therapeutic purposes Fraunhofer FIRST set up a sensor-controlled testbed. The technologies investigated include...

  • Results of measurements of ion energy losses in the range βγ ≈ 0.01-0.06 in thin absorbers. Astabatyan, R.; Lukyanov, S.; Markaryan, E.; Maslov, V.; Penionzhkevich, Yu.; Smirnov, V. // Instruments & Experimental Techniques;Mar2014, Vol. 57 Issue 2, p120 

    Results of measurements of ion energy losses in the insufficiently known range βγ = 0.01-0.06 in thin (∼2 μg/cm) absorbers are presented. For this purpose, the energy losses and time-of-flight of ions have been measured simultaneously using low-pressure avalanche gas-filled...

Share