TITLE

High density Langmuir probe array for NSTX scrape-off layer measurements under lithiated divertor conditions

AUTHOR(S)
Kallman, J.; Jaworski, M. A.; Kaita, R.; Kugel, H.; Gray, T. K.
PUB. DATE
October 2010
SOURCE
Review of Scientific Instruments;Oct2010, Vol. 81 Issue 10, p10E117
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
A high density Langmuir probe array has been developed for measurements of scrape-off layer parameters in NSTX. Relevant scale lengths for heat and particle fluxes are 1-5 cm. Transient edge plasma events can occur on a time scale of several milliseconds, and the duration of a typical plasma discharge is ∼1 s. The array consists of 99 individual electrodes arranged in three parallel radial rows to allow both swept and triple-probe operation and is mounted in a carbon tile located in the lower outer divertor of NSTX between two segments of the newly installed liquid lithium divertor. Initial swept probe results tracking the outer strike point through probe flux measurements are presented.
ACCESSION #
54858126

 

Related Articles

  • Retarding field analyzer for ion energy distribution measurements at a radio-frequency biased electrode. Gahan, D.; Dolinaj, B.; Hopkins, M. B. // Review of Scientific Instruments;Mar2008, Vol. 79 Issue 3, p033502 

    A retarding field energy analyzer designed to measure ion energy distributions impacting a radio-frequency biased electrode in a plasma discharge is examined. The analyzer is compact so that the need for differential pumping is avoided. The analyzer is designed to sit on the electrode surface,...

  • Scanning ion sensitive probe for plasma profile measurements in the boundary of the Alcator C-Mod tokamak. Brunner, D.; LaBombard, B.; Ochoukov, R.; Whyte, D. // Review of Scientific Instruments;May2013, Vol. 84 Issue 5, p053507 

    A new Ion Sensitive Probe head has been created for the outer-midplane scanning probe system on the Alcator C-Mod tokamak. The new probe head contains three elements: an ion sensitive probe to measure ion temperature and plasma potential, a Langmuir probe to measure electron temperature,...

  • Plasma properties of a DC hollow cathode discharge. Bhuyan, H.; Valderrama, E.; Favre, M.; Chuaqui, H.; Mitchell, I.; Wyndham, E. // AIP Conference Proceedings;2006, Vol. 875 Issue 1, p401 

    We have investigated the plasma properties of a Hollow Cathode Discharge (HCD) using different dimensions of hollow cathode aperture (1 to 8 mm) in argon operating medium. The diagnostic used to investigate the plasma properties is a tiny Langmuir probe. The diameter and length of the Langmuir...

  • Mirror Langmuir probe: A technique for real-time measurement of magnetized plasma conditions using a single Langmuir electrode. LaBombard, B.; Lyons, L. // Review of Scientific Instruments;Jul2007, Vol. 78 Issue 7, p073501 

    A new method for the real-time evaluation of the conditions in a magnetized plasma is described. The technique employs an electronic “mirror Langmuir probe” (MLP), constructed from bipolar rf transistors and associated high-bandwidth electronics. Utilizing a three-state bias wave...

  • A maximum-likelihood algorithm for reduction of Langmuir probe data. Martin, Adam K.; Koelfgen, Syri J. // Review of Scientific Instruments;Oct2007, Vol. 78 Issue 10, p103508 

    The reduction of Langmuir triple and quadruple probe data, i.e., the determination of the electron temperature Te from the measured voltages and currents, requires the solution of an implicit transcendental equation in Te, at every point in time. Random errors and noise in the measurements...

  • Arc discharge regulation of a megawatt hot cathode bucket ion source for the experimental advanced superconducting tokamak neutral beam injector. Xie, Yahong; Hu, Chundong; Liu, Sheng; Jiang, Caichao; Li, Jun; Liang, Lizhen; Team, NBI // Review of Scientific Instruments;Jan2012, Vol. 83 Issue 1, p013301 

    Arc discharge of a hot cathode bucket ion source tends to be unstable what attributes to the filament self-heating and energetic electrons backstreaming from the accelerator. A regulation method, which based on the ion density measurement by a Langmuir probe, is employed for stable arc discharge...

  • Electron attachment to MoF6, ReF6, and WF6; reaction of MoF6- with ReF6 and reaction of Ar+ with MoF6. Friedman, Jeffrey F.; Stevens, Amy E.; Miller, Thomas M.; Viggiano, A. A. // Journal of Chemical Physics;6/14/2006, Vol. 124 Issue 22, p224306 

    Rate constants were measured for electron attachment to MoF6, ReF6, and WF6 in 133 Pa of helium gas using a flowing-afterglow Langmuir-probe apparatus. The experiment is a thorny one because the molecules tend to form oxide impurities on feedline surfaces and because of thermal decomposition of...

  • Langmuir probe study in the nonresonant current drive regime of helicon discharge. Paul, Manash Kumar; Bora, Dhiraj // Pramana: Journal of Physics;Jul2008, Vol. 70 Issue 7, p117 

    Characterization of the current drive regime is done for helicon wave- generated plasma in a torus, at a very high operating frequency. A radiofrequency- compensated Langmuir probe is designed and used for the measurement of plasma parameters along with the electron energy distributions in...

  • Direct comparison of turbulence measurements by Langmuir probe technique and by reflectometry. Urazbaev, A.; Vershkov, V.; Soldatov, S.; Shelukhin, D. // Plasma Physics Reports;Aug2006, Vol. 32 Issue 8, p619 

    A comparative analysis of the experimental results from two diagnostic methods�reflectometry and Langmuir probe technique�is carried out, and advantages and drawbacks of these methods are considered; in particular, a comparison is made between their spectral characteristics. The...

Share

Read the Article

Courtesy of VIRGINIA BEACH PUBLIC LIBRARY AND SYSTEM

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library