Purpose. The use of continuous quality-improvement (CQI) processes in the implementation of intelligent infusion pumps in a community teaching hospital is described. Summary. After the decision was made to implement intelligent i.v. infusion pumps in a 413-bed, community teaching hospital, drug libraries for use in the safety software had to be created. Before drug libraries could be created, it was necessary to determine the epidemiology of medication use in various clinical care areas. Standardization of medication administration was performed through the CQI process, using practical knowledge of clinicians at the bedside and evidence-based drug safety parameters in the scientific literature. Post-implementation, CQI allowed refinement of clinically important safety limits while minimizing inappropriate, meaningless soft limit alerts on a few select agents. Assigning individual clinical care areas (CCAs) to individual patient care units facilitated customization of drug libraries and identification of specific CCA compliance concerns. Between June 2007 and June 2008, there were seven library updates. These involved drug additions and deletions, customization of individual CCAs, and alterations of limits. Overall compliance with safety software use rose over time, from 33% in November 2006 to over 98% in December 2009. Many potentially clinically significant dosing errors were intercepted by the safety software, prompting edits by end users. Only 4--6% of soft limit alerts resulted in edits. Conclusion. Compliance rates for use of infusion pump safety software varied among CCAs over time. Education, auditing, and refinement of drug libraries led to improved compliance in most CCAs.
ACCESSION #
53053409

 

Related Articles

  • Measures to improve safety of an elastomeric infusion system for pain management. Birrer, Kara L.; Anderson, Rebecca L.; Liu-DeRyke, Xi; Patel, Kuldip R. // American Journal of Health-System Pharmacy;7/1/2011, Vol. 68 Issue 13, p1251 

    Purpose. Measures to improve the safe implementation and utilization of an elastomeric infusion system for pain management are described. Summary. Due to the multiple safety concerns associated with the use of the On-Q infusion systems (I-Flow Corporation, Lake Forest, CA) in a community-based...

  • Smart pumps-good for nurses as well as patients. Upton, David; Quinn, Chris // British Journal of Nursing;7/25/2013 Supplement, Vol. 22 Issue S14, p4 

    Much has been made of the risks to patients, particularly from drug overdose, inherent in delivering medication via the intravenous route. The call for enhanced patient safety has fuelled the development of technological solutions to reduce the consequences of human error. However, while patient...

  • Infusion device standardisation and dose error reduction software. Iacovides, Ioanna; Blandford, Ann; Cox, Anna; Franklin, Bryony Dean; Lee, Paul; Vincent, Chris J. // British Journal of Nursing;7/24/2014 Biopatch Supplement, Vol. 23 Issue 14, pS16 

    In 2004, the National Patient Safety Agency (NPSA) released a safety alert relating to the management and use o f infusion devices in England and Wales. The alert called for the standardisation of infusion devices and a consideration o f using centralised equipment systems to manage device...

  • Safer ambulatory syringe drivers: experiences of one acute hospital trust. Freemantle, Alison; Clark, Daniel; Crosby, Vincent // International Journal of Palliative Nursing;Feb2011, Vol. 17 Issue 2, p86 

    In December 2010 the National Patient Safety Agency released a Rapid Response Report citing evidence of harm and death necessitating the need to immediately address safety issues affecting ambulatory syringe drivers with rate settings in millimetres of travel. These devices are widely used in...

  • Morphine.  // Reactions Weekly;1/5/2008, Issue 1183, p22 

    The article describes the case of a male patient who has an intraethical morphine pump. After receiving a morphine injection, the patient developed erythematous wheals. The morphine was administered into his pump. The patient was observed for possible sedation or respiratory depressions. His...

  • Medtronic recalls infusion pump linked to 14 deaths.  // Medical Device Daily;6/28/2013, Vol. 17 Issue 125, p10 

    The article reports on the announcement by medical device company Medtronic, that its SynchroMed infusion pumps have flaws which led to 14 patient deaths which prompted U.S. regulators to issue a recall for the device.

  • HTSI honored for patient safety work.  // Healthcare Purchasing News;Jan2014, Vol. 38 Issue 1, p6 

    The article announces that Healthcare Technology Safety Institute (HTSI) of AAMI Foundation was chosen as one of the recipients of the annual Cheers Awards of the Institute for Safe Medication Practices (ISMP).

  • Medication underdosing and underprescribing: Issues that may contribute to polypharmacy, poor outcomes. Halczli, Alyssa; Woolley, Adam B. // Formulary;Jun2013, Vol. 48 Issue 6, p194 

    Medication underdosing and underprescribing are often overlooked and can result in poor patient outcomes. They can also contribute to polypharmacy and significant cost to the healthcare system. Pharmacists can play a key role in preventing underdosing and underprescribing of medications by...

  • Smart pumps infuse safety into intelligent system design. Akridge, Jeannie // Healthcare Purchasing News;Apr2013, Vol. 37 Issue 4, p18 

    The article reports on topics, including patient safety, infusion devices and infusion pumps, that were discussed at the 2010 Association for the Advancement of Medical Instrumentation/Food & Drug Administration Infusion Device Summit. A discussion of several infusion devices which are suitable...