TITLE

Clinical evaluation of nine hollow-fibre membrane oxygenators

AUTHOR(S)
Segers, P.; Heida, J.; de Vries, I.; Maas, C.; Boogaart, A.; Eilander, S.
PUB. DATE
March 2001
SOURCE
Perfusion;Mar2001, Vol. 16 Issue 2, p95
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
In a comparative study we investigated the performance characteristics of nine hollow-fibre oxygenators. In a clinical setting, 10 units of each type of oxygenator were tested for oxygen exchange, transoxygenator pressure drop, heat exchanger performance and blood trauma. The oxygenators included are Maxima PRF Plus, Affinity, Forté, Affinity NT, Quantum, Optima, Capiox 1.8, Hilite and Quadrox. Ninety patients scheduled for elective coronary bypass surgery were enrolled in the study. Cardiotomy suction blood was not allowed to mix with the circulating blood. Samples were taken preoperatively (t[sub 1]), immediately after cooling and crossclamping of the aorta (t[sub 2]), during rewarming (t[sub 3]), and postoperatively (t[sub 4]). The results showed that under clinical conditions all of the tested oxygenators are well capable of meeting the normal and above-normal oxygen needs. The Quantum and Affinity NT are the best performing oxygenators with respect to oxygen transfer. The Quantum oxygenator with a small membrane surface area of 1.9 m[sup 2] has also the highest oxygen transfer per square metre. A direct relationship between shunt fraction and blood flow could not be established in all oxygenators. The lowest shunt fraction at 4 litres blood flow was seen in the Maxima, Quantum and Affinity NT groups (15%). At near physiological PaO[sub 2]-values, we noticed a negative influence on the shunt fraction. Large differences exist in pressure drop between the oxygenators under standardized conditions with circulating priming, as well as during bypass. The lowest pressure drop was noticed in the Quadrox and Affinity NT, the highest in the Quantum, Capiox and Optima. However, we did not measure any differences in pressure drop after cooling and during re-warming. Differences exist between the heat exchanger performances. The Quadrox oxygenator is the most efficient, while the Maxima has the smallest performance factor. It was not possible to calculate heat exchanger performance for the Hilite due to incorrect temperature measurement. The high pressure gradients in some of the devices did not result in higher haemolysis. Plasma free haemoglobin values were low in each oxygenator. There are no differences in platelet drop postoperatively. The influence on blood trauma of the higher pressure drop in some of the tested devices, in combination with the higher centrifugal pump revolutions needed to overcome this gradient, has to be studied with longer perfusion times.
ACCESSION #
4524369

 

Related Articles

  • A Metabolic Model of Human Erythrocytes: Practical Application of the E-Cell Simulation Environment. Yachie-Kinoshita, Ayako; Nishino, Taiko; Shimo, Hanae; Suematsu, Makoto; Tomita, Masaru // Journal of Biomedicine & Biotechnology;2010, p1 

    The human red blood cell (RBC) has long been used for modeling of complex biological networks, for elucidation of a wide variety of dynamic phenomena, and for understanding the fundamental topology of metabolic pathways. Here, we introduce our recent work on an RBC metabolic model using the...

  • Cell-free plasma hemoglobin removal by dialyzers with various permeability profiles. Hulko, Michael; Kunz, Melanie; Yildirim, Mehmet; Homeyer, Sandra; Amon, Oliver; Krause, Bernd // Scientific Reports;11/13/2015, p16367 

    The release of hemoglobin from mechanically stressed erythrocytes into plasma is a general side effect of extracorporeal therapies, such as extracorporeal membrane oxygenation or hemodialysis. In many reported cases dialysis patients showed elevated cell-free plasma hemoglobin (CPH) levels which...

  • Should critically ill patients be routinely transfused to a normal haemoglobin level? Markhorst, D. G.; Kneyber, M. C. J. // Archives of Disease in Childhood;Nov2007, Vol. 92 Issue 11, p1038 

    The article presents a case study on whether critically ill patients should be transfused to a normal hemoglobin level among children in Great Britain. Anemia is considered common among critically ill patients and may have multiple causes. It is suggested that hemoglobin concentration should be...

  • Increased hemolysis from saline pre-washing RBCs or centrifugal pumps in neonatal ECMO. Masalunga, C.; Cruz, M.; Porter, B.; Roseff, S.; Chui, B.; Mainali, E. // Journal of Perinatology;Jun2007, Vol. 27 Issue 6, p380 

    Objectives:Hemolysis is a significant complication of extracorporeal membrane oxygenation (ECMO), with a reported incidence of 12.2%. The aims of this study were (1) to investigate hemolysis caused by saline-washed versus unwashed RBCs, (2) to determine in vitro the effects of saline washing on...

  • Evaluation of the System O2 Inc portable non-pressurized oxygen delivery system [Abstract]. Pollock, N.W.; Hobbs, G.W. // Diving & Hyperbaric Medicine: Journal of the South Pacific Under;Mar2007, Vol. 37 Issue 1, p45 

    An abstract of the article "Evaluation of the System O2 Inc Portable Non-Pressurized Oxygen Delivery System," by NW Pollock and GW Hobbs is presented.

  • Prognosis in the Patients with Prolonged Extracorporeal Membrane Oxygenation. Tae-Hun Kim; Cheong Lim; Il Park; Dong Jin Kim; Yochun Jung; Kay-Hyun Park // Korean Journal of Thoracic & Cardiovascular Surgery;2012, Vol. 45 Issue 4, p236 

    Background: Prolonged usage of extracorporeal membrane oxygenation (ECMO) may induce multi-organ failure. This study is aimed to evaluate prognostic factors in the patients with ECMO. Also, the prognosis of ECMO with Kidney Injury Network Scoring system is studied. Materials and Methods: From...

  • Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation. Thiara, A. P. S.; Hoel, T. N.; Kristiansen, F.; Karlsen, H. M.; Fiane, A. E.; Svennevig, J. L. // Perfusion;Sep2007, Vol. 22 Issue 5, p323 

    Objective: Two extracorporeal membrane oxygenation (ECMO) circuits for children under 10 kg were evaluated and compared for plasma leakage, hemolysis, blood transfusions, and durability. Methods: Group A (n = 20) was supported by ECMO circuits with the Minimax oxygenator and the Biomedicus...

  • Plasma Free Hemoglobin Is an Independent Predictor of Mortality among Patients on Extracorporeal Membrane Oxygenation Support. Omar, Hesham R.; Mirsaeidi, Mehdi; Socias, Stephanie; Sprenker, Collin; Caldeira, Christiano; Camporesi, Enrico M.; Mangar, Devanand // PLoS ONE;Apr2015, Vol. 10 Issue 4, p1 

    Background: Hemolysis is common in all extracorporeal circuits as evident by the elevated plasma free hemoglobin (PFHb) level. We investigated whether increased hemolysis during extracorporeal membrane oxygenation (ECMO) is an independent mortality predictor. Methods: We performed a...

  • Oxygen delivery, carbon dioxide removal, energy transfer to lungs and pulmonary hypertension behavior during venous-venous extracorporeal membrane oxygenation support: a mathematical modeling approach. Pinheiro Besen, Bruno Adler Maccagnan; Romano, Thiago Gomes; Zigaib, Rogerio; Mendes, Pedro Vitale; Melro, Lívia Maria Garcia; Park, Marcelo // Revista Brasileira de Terapia Intensiva;Apr-Jun2019, Vol. 31 Issue 2, p113 

    Objective: To describe (1) the energy transfer from the ventilator to the lungs, (2) the match between venous-venous extracorporeal membrane oxygenation (ECMO) oxygen transfer and patient oxygen consumption (VO2), (3) carbon dioxide removal with ECMO, and (4) the potential effect of systemic...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics