TITLE

Short wavelength far infrared laser polarimeter with silicon photoelastic modulators

AUTHOR(S)
Akiyama, T.; Kawahata, K.; Tanaka, K.; Okajima, S.; Nakayama, K.
PUB. DATE
October 2008
SOURCE
Review of Scientific Instruments;Oct2008, Vol. 79 Issue 10, p10E720
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
A short wavelength far infrared laser whose wavelength λ is about 50 μm is preferable for a polarimeter and an interferometer for high density operations in the Large Helical Device (LHD) and on future large fusion devices such as ITER. This is because the beam bending effect (∝λ2) in a plasma, which causes fringe jump errors, is small and the Faraday and the Cotton–Mouton effects are moderate. We have developed a polarimeter with highly resistive silicon photoelastic modulators (PEMs) for the CH3OD laser (λ=57.2 and 47.7 μm). We performed bench tests of the polarimeter with a dual PEM and demonstrated the feasibility for the polarimeter. Good linearity between actual and evaluated polarization angles is achieved with an angular resolution of 0.05° and a temporal resolution of 1 ms. The baseline drift of the polarization angle is about 0.1° for 1000 s.
ACCESSION #
35150171

 

Related Articles

  • Multi-channel far-infrared HL-2A interferometer-polarimeter. Zhou, Y.; Deng, Z. C.; Li, Y. G.; Yi, J. // Review of Scientific Instruments;Oct2012, Vol. 83 Issue 10, p10E336 

    An HL-2A interferometer is upgraded to a multi-channel interferometer/polarimeter, which includes four chords for the interferometer and four chords for the polarimeter. The far-infrared lasers (at λ = 432.5 μm and 30 mW power) are used to probe plasmas horizontally in the midplane of...

  • Multichannel far-infrared phase imaging for fusion plasmas. Young, P. E.; Neikirk, D. P.; Tong, P. P.; Rutledge, D. B.; Luhmann, N. C. // Review of Scientific Instruments;Jan1985, Vol. 56 Issue 1, p81 

    A 20-channel far-infrared imaging interferometer system has been used to obtain single-shot density profiles in the UCLA Microtor tokamak. This system differs from conventional multichannel interferometers in that the phase distribution produced by the plasma is imaged onto a single, monolithic,...

  • Far-infrared polarimetry diagnostic for measurement of internal magnetic field dynamics and fluctuations in the C-MOD Tokamak (invited). Bergerson, W. F.; Xu, P.; Irby, J. H.; Brower, D. L.; Ding, W. X.; Marmar, E. S. // Review of Scientific Instruments;Oct2012, Vol. 83 Issue 10, p10E316 

    A laser-based (2.55 THz) mulitchord polarimeter is now operational on Alcator C-Mod and is used to make measurements of the internal magnetic field structure as well as plasma fluctuations. The polarimeter is designed to measure the Faraday effect for high-field (up to 8.3 T) and high-density...

  • Development and design of a nonmagnetic and nonconducting interferometer optical system. Burns, S. R.; Peebles, W. A.; Brooksby, C. A.; Rice, B. W. // Review of Scientific Instruments;Aug1988, Vol. 59 Issue 8, p1582 

    The use of a 0.185-mm far-infrared (FIR) laser interferometer on MTX required the design and development of optical components which are nonmagnetic and nonconducting. This arose from the requirement that the path length of the FIR interferometer not change by more than 1/50 of a wavelength...

  • Recent developments of the JET far-infrared interferometer-polarimeter diagnostic. Boboc, A.; Gelfusa, M.; Murari, A.; Gaudio, P.; JET-EFDA Contributors // Review of Scientific Instruments;Oct2010, Vol. 81 Issue 10, p10D538 

    The far-infrared diagnostic provides essential internal measurements of the plasma density and magnetic field topology (q-profile via Faraday rotation angle) in real-time. The diagnostic capabilities have recently been extended in a number of key areas. Fast interferometer data, with 10 μs...

  • High resolution polarimeter-interferometer system for fast equilibrium dynamics and MHD instability studies on Joint-TEXT tokamak (invited). Chen, J.; Zhuang, G.; Li, Q.; Liu, Y.; Gao, L.; Zhou, Y. N.; Jian, X.; Xiong, C. Y.; Wang, Z. J.; Brower, D. L.; Ding, W. X. // Review of Scientific Instruments;2014, Vol. 85 Issue 11, p1 

    A high-performance Faraday-effect polarimeter-interferometer system has been developed for the J-TEXT tokamak. This system has time response up to 1 µs, phase resolution < 0.1° and minimum spatial resolution > 15 mm. High resolution permits investigation of fast equilibrium dynamics as...

  • JET polari-interferometer. Braithwaite, G.; Gottardi, N.; Magyar, G.; O’Rourke, J.; Ryan, J.; Véron, D. // Review of Scientific Instruments;Sep89, Vol. 60 Issue 9, p2825 

    A multichannel far-infrared interferometer used on the Joint European Torus (JET) is described. The light source is a 195-µm DCN laser. The instrument is of the Mach-Zehnder type, with a heterodyne detection system. The modulation frequency (100 kHz) is produced by diffraction from a rotating...

  • Multiple view fan beam polarimetry on Tokamak devices Geck, W. R.; Domier, C. W.; Luhmann, Jr., N. C. // Review of Scientific Instruments;Jan1997, Vol. 68 Issue 1, p416 

    Describes a polarimeter diagnostic which utilizes several fan beams to accumulate line integrated Faraday rotation data in a Tokamak plasma. Magnetic expansion technique; Multiple view reconstruction method.

  • Dichroic filters to protect milliwatt far-infrared detectors from megawatt ECRH radiation. Bertschinger, G.; Endres, C. P.; Lewen, F.; Oosterbeek, J. W. // Review of Scientific Instruments;Oct2008, Vol. 79 Issue 10, p10E709 

    Dichroic filters have been used to shield effectively the far infrared (FIR) detectors at the interferometer/polarimeter on TEXTOR. The filters consist of metal foils with regular holes, the hole diameter, the mutual spacing and the thickness of the foils are chosen to transmit radiation at the...