TITLE

Rim Cracking of the Cross-Linked Longevity Polyethylene Acetabular Liner After Total Hip Arthroplasty

AUTHOR(S)
Tower, Stephen S.; Currier, John H.; Currier, Barbara H.; Lyford, Kimberly A.; van Citters, Douglas W.; Mayor, Michael B.
PUB. DATE
October 2007
SOURCE
Journal of Bone & Joint Surgery, American Volume;Oct2007, Vol. 89-A Issue 10, p2212
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
Background: Studies have suggested that cross-linked polyethylene bearings reduce wear rates from 40% to 100% compared with conventional polyethylene. However, the reduced mechanical properties of highly cross-linked polyethylene have the potential to be a limiting factor in device performance. We reviewed a series of retrieved acetabular liners with a fracture of the superior rim to assess the factors that played a role in their failure. Methods: Four Longevity acetabular bearings, which had been retrieved from two patients after seven to twenty-seven months in vivo, were visually examined for clinical damage, were assessed with use of Fourier transform infrared spectroscopy to determine the level of oxidation, and were analyzed for mechanical properties and fracture surface characterization. Control data were obtained from never-implanted devices and from global reference ultra-high molecular weight polyethylene bar stock as an industry calibration material. Results: All four retrieved liners demonstrated articular surface wear modes, which in most cases were rated as moderate, and none were rated as severe. All showed cracking or rim failure of the liner at the superior aspect along the groove in the polyethylene that engages the locking ring of the shell. The retrieved liners had no measurable oxidation, and the mechanical properties were comparable with those of never-implanted material. Conclusions: There was no notable in vivo degradation of the retrieved liners. Important factors related to failure appear to be thin polyethylene at the cup rim, relatively vertical cup alignment, and the material properties of the highly cross-linked polyethylene that are decreased relative to conventional polyethylene. The critical dimension with respect to rim failure in modular liners appears to be the minimum thickness at the equatorial region. Clinical Relevance: For a given implant design and loading, highly cross-linked polyethylene may be more susceptible to fatigue damage, such as rim cracking, than is conventional polyethylene. The potential for excess rim loading on thin polyethylene should be assessed carefully when the use of a cross-linked liner and a large femoral head is being considered, particularly in a cup with a more vertical abduction angle.
ACCESSION #
27093997

 

Related Articles

  • Correlation between in vivo stresses and oxidation of UHMWPE in total hip arthroplasty. Regis, M.; Bracco, P.; Giorgini, L.; Fusi, S.; Dalla Pria, P.; Costa, L.; Schmid, C. // Journal of Materials Science: Materials in Medicine;Sep2014, Vol. 25 Issue 9, p2185 

    The possibility of in vivo, stress-induced oxidation in orthopaedic UHMWPE has been investigated. EtO sterilised, uncrosslinked UHMWPE liners, explanted or shelf-aged, have been collected. Linear wear and wear rate were assessed and FTIR spectroscopy was employed to detect oxidation and to build...

  • Proper implant selection can quiet ceramic hips. Murphy, Stephen B. // Orthopedics Today;Apr2008, Vol. 28 Issue 4, p92 

    The article discusses a study conducted by doctor Stephen B. Murphy and other surgeons to observe the clinical results of their ceramic hip implants over an 11-year period, and revisions of ceramic hips and a 5-year period in relation to the issue of squeaking ceramic hip. Results revealed that...

  • The S-ROM Stem and Its Use With Alternative Bearings. Politi, Joel // Orthopedics;Sep2005 Supplement, Vol. 28, ps1053 

    Many long-term studies of total hip arthroplasty (THA) show excellent results. This long-term success depends on many factors, including implant fixation and bearing surface wear. As THA is commonly performed on patients with a steadily increasing life span and activity level, the issue of wear...

  • Analysis of Retrieved Acetabular Components of Three Polyethylene Types. Matias Salineros // Clinical Orthopaedics & Related Research;Dec2007, Vol. 465 Issue 1, p140 

    The polyethylene used in total hip arthroplasty has gone through many changes over the past several decades, including consolidation processes, resin types, method of sterilization, packaging, and the extent of crosslinking. To isolate the in vivo performance of material changes from implant...

  • Phase transformation of a new generation yttria-stabilized zirconia femoral head after total hip arthroplasty. Hayaishi, Yasuhisa; Miki, Hidenobu; Yoshikawa, Hideki; Sugano, Nobuhiko // Modern Rheumatology (Springer Science & Business Media B.V.);Dec2008, Vol. 18 Issue 6, p647 

    We examined a new-generation yttria-stabilized zirconia head manufactured by NGK 1 year after total hip arthroplasty. Monoclinic content of the retrieved head was twice that of the unused head at the pole and equator. A fourfold increase in monoclinic content was observed at 5 mm below the...

  • Influence of the acetabular cup position on hip load during arthroplasty in hip dysplasia. Bicanic, Goran; Delimar, Domagoj; Delimar, Marko; Pecina, Marko // International Orthopaedics;Apr2009, Vol. 33 Issue 2, p397 

    Placement of the acetabular cup during total hip arthroplasty is of great importance because usually every deviation from the ideal centre of rotation negatively influences endoprosthesis survival, polyethylene wear and hip load. Here we present hip load change in respect to various acetabular...

  • Clinical Performance of Highly Cross - Linked Polyethylenes in Total Hip Arthroplasty. Jacobs, Cale A.; Christensen, Christian P.; Greenwald, A. Seth; McKellop, Harry // Journal of Bone & Joint Surgery, American Volume;Dec2007, Vol. 89-A Issue 12, p2779 

    The article discusses the clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. Polyethylene wear of total hip prostheses is affected by multiple factors, including the type of resin used, the diameter of the femoral head and the alignment of the acetabular and...

  • Ceramic-on-Ceramic Total Hip Arthroplasty Early Dislocation Rate. Clifford Colwell // Clinical Orthopaedics & Related Research;Dec2007, Vol. 465 Issue 1, p155 

    Wear debris from metal-on-polyethylene articulation in conventional total hip arthroplasty can limit the implantʼs longevity. Modern ceramic material with high wear resistance and low fracture risk has the potential to extend the lifetime of total hip arthroplasty, which makes the procedure...

  • Nationwide investigation into adverse tissue reactions to metal debris after metal-on-metal total hip arthroplasty in Japan. Sugano, Nobuhiko; Iida, Hirokazu; Akiyama, Haruhiko; Takatori, Yoshio; Nagoya, Satoshi; Hasegawa, Masahiro; Kabata, Tamon; Hachiya, Yudo; Yasunaga, Yuji // Journal of Orthopaedic Science;Jan2014, Vol. 19 Issue 1, p85 

    Background: Adverse reactions to metal debris (ARMD) after receiving metal-on-metal (MoM) hip implants is a recent concern. However, no epidemiologic study has examined ARMD for MoM hip implants in Japan. The purposes of this study were to research the incidence of ARMD and to identify poorly...

Share

Read the Article

Courtesy of VIRGINIA BEACH PUBLIC LIBRARY AND SYSTEM

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics