ABSTRACT
Background: Recent studies have suggested that cemented femoral components with a polished surface may pro- vide superior long-term fixation when compared with femoral components with a roughened surface. The purpose of this study was to evaluate the results of total hip arthroplasty with a cemented femoral component with a polished surface finish and compare them with the results of total hip arthroplasty performed with a similar design of cemented femoral component with a rougher surface finish. Methods: We retrospectively reviewed a consecutive group of 132 patients (149 hips) in whom primary total hip arthroplasty had been performed by one surgeon using a cemented collared femoral component with a polished (0.1-µm Ra) surface finish and a cementless acetabular component. Ninety-eight patients (115 hips) were followed for a minimum of ten years. We compared the survivorship of this prosthesis with that of a femoral component of similar design but with rougher surfaces (matte or grit-blasted). Results: No polished stems were revised because of aseptic loosening or demonstrated radiographic evidence of loosening; however, eight hips (5.4%) with a polished stem demonstrated osteolysis distal to the greater or lesser trochanter. In contrast, six stems (2.0%) with a matte surface finish of 0.8-µm Ra were revised because of aseptic loosening, and an additional five stems were seen to be loose radiographically. Eleven stems (9.2%) with a grit-blasted surface finish of 2.1-µm Ra were revised because of aseptic loosening, and an additional four stems were seen to be loose radiographically. The difference in the prevalence of revision due to aseptic loosening between the group with the 0.1-µm Ra surface and the group with the 2.1-µm Ra surface was significant (p = 0.001), as was the difference between the prevalence of revision due to aseptic loosening between the group with the 0.8-µm Ra surface and the group with the 2.1-µm Ra surface (p = 0.001). No cups were revised because of aseptic loosening, and one hip had radiographic signs of acetabular loosening. Conclusions: This study demonstrated excellent durability of a prosthesis consisting of a cemented, collared, polished femoral component and a cementless acetabular component. While no hips were revised because of aseptic loosening, distal femoral osteolysis was observed in eight hips (5.4%), a higher prevalence than has been reported by others after similar durations of follow-up of tapered, collarless, polished femoral components.
ACCESSION #
23693807

 

Related Articles

  • METALLOSIS AFTER CONTEMPORARY METAL-ON-METAL TOTAL HIP ARTHROPLASTY. Korovessis, P.; Petsinis, G.; Repanti, M.; Repantis, T. // Journal of Bone & Joint Surgery, American Volume;Jun2006, Vol. 88-A Issue 6, p1183 

    Background: Authors of recent studies have reported early periprosthetic osteolysis in patients who have been treated with a contemporary metal-on-metal total hip arthroplasty and have suggested that metal hypersensitivity associated with an immunologic response to metal may be of etiologic...

  • Hip resurfacing in patients under thirty years old: an attractive option for young and active patients. Krantz, Nicolas; Miletic, Bruno; Migaud, Henri; Girard, Julien // International Orthopaedics;Sep2012, Vol. 36 Issue 9, p1789 

    Purpose: Metal-on-metal hip resurfacing is offered as an alternative to traditional hip arthroplasty for young, active adults with advanced osteoarthritis. The concept of hip resurfacing is considered very attractive for this specific population (hard-on-hard bearing component with a large...

  • Alumina-on-alumina total hip replacement for femoral neck fracture in healthy patients. Solarino, Giuseppe; Piazzolla, Andrea; Mori, Claudio M.; Moretti, Lorenzo; Patella, Silvio; Notarnicola, Angela // BMC Musculoskeletal Disorders;2011, Vol. 12 Issue 1, p32 

    Background: Total hip replacement is considered the best option for treatment of displaced intracapsular fractures of the femoral neck (FFN). The size of the femoral head is an important factor that influences the outcome of a total hip arthroplasty (THA): implants with a 28 mm femoral head are...

  • Survival of Polished Compared with Precoated Roughened Cemented Femoral Components. Lachiewicz, Paul F.; Kelley, Scott S.; Soileau, Elizabeth S. // Journal of Bone & Joint Surgery, American Volume;Jul2008, Vol. 90-A Issue 7, p1457 

    Background: The optimal surface finish for cemented femoral components remains controversial. The purpose of this randomized clinical trial was to compare the survival of two femoral components with similar geometry but substantially different surface finishes. Methods: During a five-year...

  • Cement mantle defects in total hip arthroplasty: influence of stem size and cementing technique. A. Katzer; A. Ince; M. Hahn; M. Morlock; W. Steens // Journal of Orthopaedics & Traumatology;Dec2007, Vol. 8 Issue 4, p167 

    Abstract  The cause of isolated osteolysis in the femoral shaft around stem implants in patients with cemented THR has so far not been established. A number of factors have been considered such as torsional stability of the femoral stem implant, the time of reduction...

  • The Biology of Aseptic Osteolysis. G Holt // Clinical Orthopaedics & Related Research;Jul2007, Vol. 460 Issue 1, p240 

    Total hip arthroplasty is one of the most commonly performed and successful elective orthopaedic procedures. However, numerous failure mechanisms limit the long-term success including aseptic osteolysis, aseptic loosening, infection, and implant instability. Aseptic osteolysis and subsequent...

  • Retention of a well-fixed acetabular component in the setting of acetabular osteolysis. Kang, Pengde; Yang, Jing; Zhou, Zongke; Shen, Bin; Pei, Fuxing // International Orthopaedics;May2012, Vol. 36 Issue 5, p949 

    Purpose: The treatment strategy for pelvic osteolysis with a well-fixed acetabular component after total hip arthroplasty (THA) involves replacing the acetabular cup liner and femoral head, débriding osteolytic lesions, and grafting. Methods: We investigated whether retention of a well-fixed...

  • High survival in young patients using a second generation uncemented total hip replacement. Streit, Marcus; Schröder, Kerstin; Körber, Matthias; Merle, Christian; Gotterbarm, Tobias; Ewerbeck, Volker; Aldinger, Peter // International Orthopaedics;Jun2012, Vol. 36 Issue 6, p1129 

    Purpose: Whilst excellent long-term results with contemporary uncemented stems have been reported for total hip arthroplasty in young patients, the survival rates for the whole reconstruction are often compromised by high failure rates on the acetabular site due to peri-acetabular osteolysis and...

  • Seventeen-year survival of the cementless CLS Spotorno stem. Müller, Lutz Arne; Wenger, Nina; Schramm, Michael; Hohmann, Dieter; Forst, Raimund; Carl, Hans-Dieter // Archives of Orthopaedic & Trauma Surgery;Feb2010, Vol. 130 Issue 2, p269 

    For primary uncemented hip arthroplasty, various stem designs are available. The cementless CLS® Spotorno stem has been used for more than 20 years. We re-evaluated a group of patients previously examined for a 10-year follow-up to assess the clinical and radiological stem performance in the...

Share

Read the Article

Courtesy of VIRGINIA BEACH PUBLIC LIBRARY AND SYSTEM

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics