TITLE

Lipid peroxidation during initiation of extracorporeal membrane oxygenation after hypoxia in endotoxemic rabbits

AUTHOR(S)
Trittenwein, G.; Rotta, A.T.; Gunnarsson, B.; Steinhorn, D.M.
PUB. DATE
January 1999
SOURCE
Perfusion;1999, Vol. 14 Issue 1, p49
SOURCE TYPE
Academic Journal
DOC. TYPE
Article
ABSTRACT
Initiation of extracorporeal membrane oxygenation (ECMO) in septic children with severe respiratory failure often improves oxygenation but not pulmonary function. The factors affecting pulmonary function following onset of ECMO are not completely understood, but are thought to involve injury mediated, in part, by reactive oxygen species. We hypothesized that induction of ECMO using 100% oxygen as the sweep gas through the oxygenator would increase lipid peroxidation in endotoxin-primed animals after severe hypoxia. We further speculated that provision of oxygenated blood to the pulmonary circulation via venovenous ECMO would promote a greater degree of oxidative damage to the lung as compared to venoarterial ECMO. Eighteen New Zealand White rabbits were assigned to a control group (control) or two intervention groups subjected to 60 min of venoarterial or venovenous ECMO. ECMO was initiated following an intravenous challenge with 0.5 mg/kg of E. coli endotoxin and a period of global hypoxia leading to an arterial pH of 6.99 ± 0.09, PaCO[sub 2] of 10[sup 3] ± 31 mmHg and PaO[sub 2] of 27 ± 5 mmHg. Malondialdehyde (MDA), a marker of lipid peroxidation, was measured in lung tissue homogenates and in arterial plasma. Lung tissue MDA demonstrated a strong trend towards an increase in the venoarterial group (1884 ± 945 nmol/g protein) and in the venovenous group (1905 ± 758 nmol/g protein) in comparison to the control group (644 ± 71 nmol/g protein) (p = 0.1; significance at 95% in Scheffe test). Lung tissue MDA in the venovenous group had a significant correlation with mean PaO2 during ECMO by regression analysis (r[sup 2] = 0.678, p = 0.044). The change in blood MDA concentration between pre-ECMO and post-ECMO values was greater in the venovenous group (pre 1.62 ± 0.61 versus post 5.12 ± 0.2.07 mmol/l, p = 0.043) compared with that seen in the venoarterial group (pre 1.46 ± 0.38 versus post 3.9 ± 0.93 mmol/l). Our data support the hypothesis that initiation of ECMO with a circuit gas oxygen concentration of 100% after global hypoxia enhances oxidative damage to lipids in endotoxin-challenged animals. During venovenous ECMO this finding is dependent on PaO[sub 2].
ACCESSION #
1587750

 

Related Articles

  • Science commentary: Extracorporeal membrane oxygenation. Berger, Abi // BMJ: British Medical Journal (International Edition);10/03/98, Vol. 317 Issue 7163, p916 

    Focuses on the use of the extracorporeal membrane oxygenation technique which oxidized blood. Definition of extracorporeal membrane oxygenation; Number of extracorporeal membrane oxygenation centres in the United Kingdom; Insight into the use of the procedure on babies; Details on the benefits...

  • Identifying at Risk Infants Following Neonatal Extracorporeal Membrane Oxygenation. Kumar, Praveen; Shankaran, Seetha; Bedard, Mary P; Delaney-Black, Virginia // Journal of Perinatology;Jul99, Vol. 19 Issue 5, p367 

    OBJECTIVE: To identify infants at risk of death and abnormal neurodevelopmental outcome following extracorporeal membrane oxygenation (ECMO) in the neonatal period. METHODS: The medical records of 82 neonates treated with ECMO were reviewed to evaluate risk of death. All survivors were...

  • Extracorporeal membrane oxygenation saves lives. Boulton, Alison // BMJ: British Medical Journal (International Edition);7/20/96, Vol. 313 Issue 7050, p135 

    Reports on the use of extracorporeal membrane oxygenation for severely ill newborn babies. Cost of less invasive treatments; Decrease of risk of death from respiratory failure.

  • Hemodynamic troubleshooting for mechanical malfunction of the extracorporeal membrane oxygenation systems using the PPP triad of variables. Faulkner, Sherry C.; Johnson, Charles E.; Tucker, Juan L.; Schmitz, Michael L.; Fasules, James W.; Drummond-Webb, Jonathan J. // Perfusion;Sep2003, Vol. 18 Issue 5, p295 

    Prolonged usage of disposable extracorporeal membrane oxygenation (ECMO) circuitry increases the risk of mechanical complications due to breakdown or malposition of the circuit elements. Often, such complications are life threatening for the critically ill patient. Such problems need to be...

  • EXTRACORPOREAL MEMBRANE OXYGENATION.  // CONNECT: The World of Critical Care Nursing;2012, Vol. 9 Issue 1, p17 

    An abstract of the article "Extracorporeal Membrane Oxygenation," by Žiga Vrhovnik and Zoran Topić is presented.

  • Improved oxygenation with reduced recirculation during venovenous ECMO: comparison of two catheters. Rais-Bahrami, Khodayar; Walton, Dawn M.; Sell, Jeffrey E.; Rivera, Oswaldo; Mikesell, Gerald T.; Short, Billie L. // Perfusion;Nov2002, Vol. 17 Issue 6, p415 

    Objectives: To determine whether the new double-lumen catheter made by OriGen Biomedical (Austin, TX) for venovenous (VV) extracorporeal membrane oxygenation (ECMO) would reduce recirculation and improve oxygenation during VV ECMO when compared with the Kendall double-lumen catheter (Kendall...

  • The first succesful extracorporeal membranous oxygenation treatment in a child with refractory fulminant myocarditis in Turkey. Anıl, Ayşe Berna; Can, Fulya Kamit; Turhan, Soysal; Zengin, Neslihan; Anıl, Murat; Bakiler, Ali Rahmi; Doğrusöz, Buket // Anatolian Journal of Cardiology / Anadolu Kardiyoloji Dergisi;Dec2015, Vol. 15 Issue 12, p1036 

    A letter to the editor is presented in response to the article related to extracorporeal membranous oxygenation treatment of a pediatric patients presenting with acute fulminant myocarditis which was published in the previous issue of the periodical.

  • Portable Device Replaces Ventilator.  // USA Today Magazine;Feb2010, Vol. 138 Issue 2777, p14 

    The article reports on a new portable extracorporeal membrane oxygenation (ECMO) device, also known as an artificial lung, which oxygenates a patient's blood directly through a two-way connection via a single tube.

  • A paediatric ECMO case of plasma leakage through a polymethylpentene oxygenator. Gill, M. C.; O’Shaughnessy, K.; Dittmer, J. // Perfusion;Oct2015, Vol. 30 Issue 7, p600 

    Polymethylpentene (PMP) oxygenators, utilised for ECMO, are commonly believed to be resistant to plasma leakage. Whilst uncommon, plasma leakage has been previously reported with PMP fibres, both in vivo and in vitro. We describe a paediatric ECMO case during which plasma leakage occurred and...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics