Properties of a High-Temperature Platinum Resistance Thermometer up to 1350 °C

AUTHOR(S)
Arai, Masaru; Kawata, Akiyoshi
PUB. DATE
September 2003
SOURCE
AIP Conference Proceedings;2003, Vol. 684 Issue 1, p381
SOURCE TYPE
Conference Proceeding
DOC. TYPE
Article
ABSTRACT
Properties of high-temperature platinum resistance thermometers developed at NMIJ are studied in the range from 1100 °C to 1350 °C. The HTPRTs have shown drift rates of 0.1 mK/h at 1100 °C and 0.5 mK/h at 1170 °C. The W(29.7646 °C) values of the HTPRTs increased after exposure at 1100 °C, while they slightly decreased at 1150 °C and 1170 °C. After exposed at 1170 °C for 500 hours, three of six HTPRTs showed W(29.7646 °C) values larger than 1.118 07. It was also found that the decrease in resistance continued above 1200 °C up to 1350 °C. © 2003 American Institute of Physics
ACCESSION #
11188786

 

Related Articles

  • Comparison of Fixed Point Realisations between Inmetro and PTB. Santiago, J. F. N.; Petkovic, S. G.; Teixeira, R. N.; Noatsch, U.; Thiele-Krivoj, B. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p849 

    An interlaboratory comparison in the temperature range between -190 °C and 420 °C was organised between the National Institute of Quality, Normalisation and Industrial Quality (Inmetro), Brazil, and the Physikalisch Technische Bundesanstalt (PTB), Germany. This comparison followed the same...

  • An Independent Quality Check of Secondary Platinum Resistance Thermometer Calibrations above 0 °C. Ainsworth, Ronald E. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p423 

    Traditionally the use of least squares to evaluate the fit and consistency of a platinum resistance thermometer calibration requires the use of several redundant calibration points each of which adds time and expense to the calibration process. However the behavior of some secondary platinum...

  • A Simplified Method for Calibration of PRTs Used in Heat Meters. Moiseeva, N. P.; Petrov, D.V.; Karzhavin, A.V. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p417 

    Many thousands of industrial platinum resistance thermometers are used in heat meters that are devices for measurement of the heat being spent in heat-exchange water systems. That is why it is important to develop an inexpensive, quick and simple method for calibration of the IPRTs. There are...

  • Efficient Calibration Methods for Platinum Resistance Thermometers. Phillips, Richard W. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p411 

    Calibration methods for platinum resistance thermometers are simplified by reducing the number of calibration points. For applications that do not require the low uncertainties of the ITS-90 definitions, several modifications are available to improve calibration efficiency. The proposed schemes...

  • Measurement of the Insulation Resistance for the Development of High Temperature Platinum Resistance Thermometers with a Guard Electrode. Yamazawa, Kazuaki; Arai, Masaru // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p363 

    In the present study, the insulation resistance of insulators within standard platinum resistance thermometers (SPRT) is measured to evaluate the bias in temperature measurements caused by the leakage current flowing through the insulator, and a prototype SPRT to overcome these problems was...

  • Platinum-Sheathed High-Temperature Platinum Resistance Thermometers for Use up to 1200 °C. Arai, Masaru; Kawata, Akiyoshi; Imamura, Tomosuke; Kinoshita, Kazuhiro // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p357 

    A new type of High-Temperature Platinum Resistance Thermometer (HTPRT) with a platinum sheath was developed. The stability of the resistance at temperatures above 1100 °C up to 1170 °C was studied. The HTPRTs have shown resistance changes of 1.0 mK/h at 1100 °C, 0.7 mK/h at 1150 °C...

  • Oxidation of Platinum Resistance Thermometers. Ancsin, John // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p345 

    PRTs were soaked at different temperatures, and the resulting changes caused by oxidation and the subsequent dissociation of the formed oxide were observed. Soaking below 400 °C initiated oxidation, in some platinum resistance thermometers (PRTs), and then their resistance steadily increased...

  • Experimental Study of Different Filling Gases on the Stability of Metal-Sheathed Standard Platinum Resistance Thermometers. Zhao, M.; Li, X.; Chen, D. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p339 

    There is a trade-off between the oxidation effect and element contamination in metal-sheathed standard platinum resistance thermometers (MSPRTs). Excessively high O2 partial pressure causes the platinum sensor to oxidize, and excessively low O2 leads to sensor contamination. The oxygen content...

  • Performance Assessment of Resistance Ratio Bridges Used for the Calibration of SPRTs. Strouse, Gregory F.; Hill, Kenneth D. // AIP Conference Proceedings;2003, Vol. 684 Issue 1, p327 

    Automatic balancing dc and ac resistance ratio bridges are the primary measurement devices used by National Measurement Institutes for the calibration of standard platinum resistance thermometers (SPRTs) on the International Temperature Scale of 1990 (ITS-90). Performance assessment of these...

Share

Read the Article

Courtesy of THE LIBRARY OF VIRGINIA

Sorry, but this item is not currently available from your library.

Try another library?
Sign out of this library

Other Topics